Providers to have confidence in reliable infusion pump performance, the ability The new CADD®-Solis VIP system will allow health care Infusion delivery history for quick and easy therapy assessment and continuous

#CADD SOLIS PCA PUMP PC#

Their PC computer into the pump and also view reports on patient data and

#CADD SOLIS PCA PUMP SOFTWARE#

Safety software allows clinicians to download therapy information directly from Patient-focused user interface displays therapy information in anĮasy-to-understand format to help reduce training time, and the medication Hydration, pain management, and IVIG (Intravenous Immunoglobulin or when a blood product is administered intravenously) for adult and pediatric patients. Variety of therapies including TPN (Total Parenteral Nutrition), chemotherapy, The CADD®-Solis VIP system includes five delivery modes:Ĭontinuous, Intermittent, Taper, PCA and Step, which allow it to be used for a Providers who want to provide optimal care for their infusion patients in VIP system provides enhanced safety features and convenience for health care “Until now, however, Smart infusion systems have not been designedįor home patients or the professionals who care for them. Improve efficiencies,” said Tommy Johns, Vice President Global Product Hospitals as a means to help reduce the risk of medication delivery errors and “Smart infusion devices have been widely adopted in Incorporate homecare-desired features for safety, ease of use, mobility, and However, the new CADD®-Solis VIP system was adapted to The Smart CADD®-Solis hospital pain management system that Smiths Medical The new system leverages the same software and hardware features of Safety software, and dedicated medication cassette reservoirs that attach to Multi-therapy infusion platform that includes a Smart infusion pump, medication The CADD®-Solis VIP (variable infusion profile) system is a The new system is already available in Canada, Leading global medical device manufacturer, announced today it launched theĬADD®-Solis VIP ambulatory infusion system with CADD™-Solis medication safety PAUL, Minn.–( BUSINESS WIRE)–Smiths Medical, a The pump log was reviewed for unusual log/error messages and was not found to contain any messages that would indicate that an equipment malfunction occurred.ST. The pump performed within manufacturer's specifications on all tests.

#CADD SOLIS PCA PUMP FULL#

The pump was sequestered and full functional testing was performed by clinical engineering. The pump was turned off 1 hour and 45 minutes after the infusion started, and the pt died 2 hours after turning off the pump.Ī direct association between the pt's death and the medication error cannot be clearly identified. Thus, the dose of morphine delivered was five times higher than the ordered dose. The nurse then attached the medication cassette containing morphine 5 mg/ml to the pump, and started the infusion. She programmed the pca pump to deliver the correct basal rate (2 mg/hr) and incremental dose (0.ĥ mg), but because the pump was programmed as hydromorphone 1 mg/ml, the infusion rate calculated as 2 ml/hr and incremental dose as 0. However, the pca had been previously programmed to deliver hydromorphone 1 mg/ml and the nurse failed to clear the pump (by starting a new patient on the pump) before she began programming it. The palliative care team was providing end-of-life palliative treatment and a pca was ordered for morphine 5mg/ml, infuse at a basal rate of 2mg/hr and an incremental dose of 0. Previous patient's drug name and drug concentration were not cleared).ĭetails: a pt was admitted for treatment related to stage iv metastatic ovarian cancer, chronic right heart failure, severe pulmonary arterial hypertension, chronic right-sided malignant pleural effusion, acute kidney injury, hyponatremia, hydronephrosis, hypoxemia and worsening shortness of breath. Summary: a patient received 5 times the ordered dose of morphine for 1 hour anf 45 minutes due to ders pump not being completely programmed correctly (i.














Cadd solis pca pump